Bypass Filters and Homogenizers

When detecting DO, an environmental engineer can use a probe and insert it directly into the process or take a sample and deliver it to a DO analyzer. The probe eliminates the dead time of sample transportation, which degrades closed-loop control. However, the probe requires effective and unattended cleaning attachments. The advantage of sampling is that loop components are more accessible. Therefore, DO probes are used for online control, and sampling systems are used only for monitoring.

In DO detection, the measured parameter is in the liquid phase and the solids can be filtered out. However, in BOD measurement, oxygen demand occurs mostly in the solids. Therefore, they must be included in the evaluation.

Consequently, most BOD and COD analyzers require a sampling system in which homogenization (liquification of solids), is used instead of filters, to prevent plugging and retain sample integrity.